edge® digital pH meter for wine kit delivers precision while offering simplicity and versatility in a sleek design.

Upgrade to the Hanna edge® wine pH kit for professional winemakers. The HI2020W allows you to add a conductivity probe and a dissolved oxygen probe to give you full versatility. Fast results, ultra-accurate, hassle-free analysis.

  • Complete meter kit for measuring pH in wine.
  • CAL Check™ feature ensures the accuracy of the meter.
  • Great for professional and home winemakers.

Description

The Importance of pH in Wine Making

The pH of wine is important to determine because it will affect the quality of the final product in terms of taste, color, oxidation, chemical stability and other factors. Generally in winemaking, the higher the pH reading, the lower amount of acidity in the wine. Three important factors in determining the pH of wine include the ratio of malic acid to tartaric acid, the amount of potassium, and the total amount of acid present. Most wines optimally have a pH between 2.9 and 4.0, with values differing based on the type of wine. Values above pH 4.0 indicate that the wine may spoil quickly and be chemically unstable. Lower pH values allow the wine to stay fresher for a longer period and retain its original color and flavor. High pH wine is more likely to breed bacteria and become unsuitable to drink. For finished white wines, the ideal pH is between pH 3.00 and pH 3.30, while the final pH for red wine is ideally between pH 3.40 and pH 3.50. The optimal pH before the fermentation process is between pH 2.9 and pH 4.0. The pH of wine therefore not only affects the color of wine, but also the oxidation, yeast fermentation, protein stability and bacteria growth and fermentation.

The edge HI2020W wine pH meter kit includes all the necessary supplies to determine the pH of wine must and juice accurately. The meter supplied with the HI2020W is the multiparameter edge, HI2020. The HI2020 is an advanced meter that is rich in features including using digital electrodes that can be swapped out anytime to switch to a different parameter, built in rechargeable battery, two USB ports, datalogging, automatic calibration, GLP (good laboratory practice) data for last calibrations, and Hanna’s exclusive calibration check feature that will alert the user to potential problems during calibration including when to clean the electrode and when the buffer might be contaminated. The HI2020 is a versatile pH meter for winemaking that be used as a benchtop, portable and even a wall mount to conserve space. The HI2020 is one of the most advanced meters available but  the most important aspect in measuring pH is having the correct sensor for the application.

The HI2020W wine kit is packaged with the HI10480 in place of the normally supplied HI11310. The HI10480 is a specially designed pH electrode that is resistant to clogging to the solids found in grape juice and must. This double junction, refillable pH electrode has a unique ground glass junction and PTFE sleeve to provide unsurpassed performance for the wine industry.  A standard pH electrode will have a ceramic junction that can be easily clogged by solids. When this happens the pH readings will be erratic and drift over time before stabilizing on the actual pH value of the wine sample. The HI10480 with CPS technology will prevent the junction from clogging. The HI10480 is a digital electrode that when used with the edge HI2020 will change the default buffer set from pH 4.01and  7.01 to  pH 3.00 and 7.01.

The edge HI2020W wine pH meter (kit) includes the HI180-1 compact magnetic mini-stirrer. The use of a magnetic stirrer allows for proper mixing and the sample to be homogenous in consistency and temperature. The stirring of the sample also helps speed up the response time of a sensor in achieving a stable reading. 

The edge HI2020W wine pH kit also includes all the solutions for calibration and maintenance. In order to ensure the highest of accuracy in measurement it is important to bracket the calibration buffers around the expected reading. Since the pH of wine is less than pH 4.01 it is important to calibrate at a lower pH value. For calibration the kit includes the HI5003 pH 3.00 and the HI5007 pH 7.01 technical calibration buffer. Each bottle of buffer NIST traceable is supplied with a certificate of analysis that includes lot number, expiration date and reference materials used in production. 

For maintenance the The HI70635L cleaning solution for wine deposits and the HI70300L electrode storage solution. When the “clean electrode” message is displayed during the calibration of the HI2020 edge or there is a poor probe condition (i.e. less than 30%) then soak the HI10480 in the cleaning solution for 15 minutes to clean any deposits from the glass surface. After cleaning the probe should be rinsed with purified water and then allowed to soak in the storage solution for 2-3 hours. The electrode storage solution is also used when the electrode is not in use. Either place a small amount into the storage cap or into a beaker in order to maintain a properly hydrated electrode.

Additional information

Weight 2.78 kg
Parameter

Conductivity/TDS, Dissolved, Oxygen, pH

pH Accuracy

±0.002 pH, ±0.01 pH, ±0.2 mV

pH Calibration

10.01, 12.45, 4.01 (3.00†), 5 points (Standard mode) 1.68, 6.86, 7.01, 9.18, and two custom buffers; 3 points (basic mode) 4.01; 6.86; 7.01; 9.18; 10.01

pH Range

-2.00 to 16.00 pH, -2.000 to 16.000 pH, ±1000 mV

pH Resolution

0.001 pH, 0.01 pH, 0.1 mV

EC Accuracy

±1% of reading (±0.05 µS/cm or 1 digit, whichever is greater)

EC Calibration

1 point offset calibration (0.00 µS/cm in air), 1 point slope calibration in EC standard 84 µS/cm, 12.88 mS/cm, 1413 µS/cm, 5.00 mS/cm, 80.0 mS/cm and 111.8 mS/cm

EC Range

0.00 to 29.99 µS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, 30.0 to 299.9 µS/cm, 300 to 2999 µS/cm, up to 500.0 mS/cm (absolute EC)**

EC Resolution

0.01 mS/cm, 0.01 µS/cm, 0.1 mS/cm, 0.1 µS/cm, 1 µS/cm

TDS Accuracy

±1% of reading (±0.03 ppm or 1 digit, whichever is greater)

TDS Calibration

through EC calibration

TDS Range

0.00 to 14.99 ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L, 15.0 to 149.9 ppm (mg/L), 150 to 1499 ppm (mg/L), up to 400.0 g/L (absolute TDS)**, with 0.80 conversion factor

TDS Resolution

0.01 g/L, 0.01 ppm, 0.1 g/L, 0.1 ppm, 1 ppm

Temperature Accuracy

±0.5 °C, ±0.9 °F

Temperature Range

-20.0 to 120.0 ºC, -4.0 to 248.0 °F

Specifications

SKU HI2020W
Name ® Wine pH Meter (kit) – HI2020W
pH Range -2.000 to 16.000 pH, -2.00 to 16.00 pH, ±1000 mV
pH Resolution 0.001 pH, 0.01 pH, 0.1 mV
pH Accuracy ±0.002 pH, ±0.01 pH, ±0.2 mV
pH Calibration 5 points (Standard mode) 1.68, 4.01 (3.00†), 6.86, 7.01, 9.18, 10.01, 12.45, and two custom buffers; 3 points (Basic mode) 4.01; 6.86; 7.01; 9.18; 10.01
pH Temperature Compensation ATC: -5.0 to 100.0ºC; 23.0 to 212.0°F*
pH CAL Check (electrode diagnostics) Glass and reference junction diagnostics ( HI 11311 & HI 12301 only), out of calibration range , probe condition, response time
EC Range 0.00 to 29.99 µS/cm, 30.0 to 299.9 µS/cm, 300 to 2999 µS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, up to 500.0 mS/cm (absolute EC)**
EC Resolution 0.01 µS/cm, 0.1 µS/cm, 1 µS/cm, 0.01 mS/cm, 0.1 mS/cm
EC Accuracy ±1% of reading (±0.05 µS/cm or 1 digit, whichever is greater)
EC Calibration 1 point offset calibration (0.00 µS/cm in air), 1 point slope calibration in EC standard 84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm, 80.0 mS/cm and 111.8 mS/cm
TDS Range 0.00 to 14.99 ppm (mg/L), 15.0 to 149.9 ppm (mg/L), 150 to 1499 ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L, up to 400.0 g/L (absolute TDS)**, with 0.80 conversion factor
TDS Resolution 0.01 ppm, 0.1 ppm, 1 ppm, 0.01 g/L, 0.1 g/L
TDS Accuracy ±1% of reading (±0.03 ppm or 1 digit, whichever is greater)
TDS Calibration through EC calibration
EC/TDS Temperature Compensation ATC (0.0 to 100.0ºC; 32.0 to 212.0 °F), NoTC
EC to TDS Conversion Factor 0.40 to 0.80
Salinity (% NaCl) Range 0.0 to 400.0% NaCl***
Salinity (% NaCl) Resolution 0.1% NaCl
Salinity (% NaCl) Accuracy ±1% of reading (@25ºC/77ºF)
Salinity (% NaCl) Calibration 1 point with HI 7037L 100% NaCl sea water standard
Salinity (PSU) Range 2.00 to 42.00 PSU***
Salinity (PSU) Resolution 0.01 PSU
Salinity (PSU) Accuracy ±1% of reading (@25ºC/77ºF)
Salinity (PSU) Calibration through EC calibration
Salinity Range 0.0 to 80.0 g/L***
Salinity Resolution (ppt) 0.01 g/L
Salinity Accuracy (ppt) ±1% of reading (@25ºC/77ºF)
Salinity Calibration (ppt) through EC calibration
Dissolved Oxygen Range 0.00 to 45.00 ppm (mg/L), 0.0 up to 300.0% saturation
Dissolved Oxygen Resolution 0.01 ppm, 0.1% saturation
Dissolved Oxygen Accuracy ±1 digit, ±1.5% of reading
Dissolved Oxygen Calibration one or two points at 0% (HI 7040 solution) and 100% (in air)
Dissolved Oxygen Temperature Compensation ATC: 0 to 50 ºC; 32.0 to 122.0 °F
Salinity Compensation 0 to 40 g/L (with 1 g/L resolution)
Altitude Compensation -500 to 4000 m (with 100 m resolution)
Temperature Range -20.0 to 120.0 ºC, -4.0 to 248.0 °F
Temperature Resolution 0.1 °C, 0.1 °F
Temperature Accuracy ±0.5 °C, ±0.9 °F
pH Electrode HI11310 pH electrode (included)
Logging Type Manual log-on-demand (Max. 200 logs); Manual log-on-stability (Max. 200 logs); Interval logging *** (Max. 600 samples; 100 lots)
Logging Memory up to 1000*** records
Connectivity 1 micro USB port for charging and PC connectivity, 1 USB port for storage
Battery Type/Life Built-in rechargeable battery with up to 8 hours of continuous use
Power Supply 5 VDC adapter (included)
Environment 0 to 50°C (32 to 122°F), RH max 95% non-condensing
Dimensions 202 x 140 x 12.7mm (8” x 5.5” x 0.5”)
Weight 250 g (8.82 oz)

Ordering Info

HI2020W includes: USB cable, benchtop docking station with electrode holder, wall mount cradle, 5 VDC power adapter, instruction manual and quality certificate. HI10480 digital double junction, refillable with 3.5M KCl electrolyte solution pH electrode with PTFE sleeve, built in temperature sensor, and 3.5 mm connector. HI180-1 115 VAC black compact magnetic mini stirrer with Speedsafe™ technology with micro-stir bar (1). HI5003 pH 3.00 technical calibration buffer (500 ml). HI5007 pH 7.01 technical calibration buffer (500 ml). HI70635L cleaning solution for wine deposits (500 ml). HI70300L electrode storage solution (500 ml).