Description
HI97705-11 is a high quality set of calibration standards that allows users to validate and calibrate the HI96705 Silica Portable Photometer. The HI97705-11 includes one sealed “A” cuvette for zeroing the meter and one “B” cuvette that is used to validate or calibrate. The standard value of the “B” cuvette is 1.00 ±0.01 mg/L (ppm) SiO2. An acceptable reading when validating the HI97705 with the standards is 1.00 ±0.05 mg/L @ 25℃. The calibration set is produced in our state-of-the-art manufacturing facility using reagent grade chemicals in a temperature controlled environment.
Supplied with Certificate of Analysis
- Product code
- Lot number
- Expiration date
- Standard value @ 25℃
- Reference meter NIST traceable
Provided storage containers
- Light tight
- Protects from accidental breakage
Validation and calibration of the HI97705 Silica Portable Photometer is quick and easy with the HI97705-11 CAL Check™ Standards.