Hanna Instruments Canada
The Hanna Advantage: A Worldwide Leader in Technology and Innovation 100% Quality  360 Value
Contact Hanna: (800) 842-6629
Hanna Instruments Canada
Customer Support: (800) 842-6629

Ultra High Range Chlorine Portable Photometer - HI96771-kit

The HI96771 Ultra High Range Chlorine Portable Photometer combines accuracy and ease of use in an ergonomic, portable design. A user can accurately determine the concentration of chlorine over two ranges. The HI96771 has a 0.00 to 5.00 mg/L (ppm) free chlorine low range and a high range mode that measures chlorine up to 500 mg/L (ppm). The HI93701-0 reagents are used for the low range free chlorine measurement while the HI95771-0 are used for high range chlorine measurements. The HI96771 offers many advanced features including the exclusive CAL Check™ function used for performance verification and calibration of the meter.

CAL Check™ Validation and Calibration

Up to 500 ppm chlorine range

Built-in Timer

Description

The HI96771 portable photometer is for the measurement chlorine over a very wide range. Most chlorine photometers are limited to very low concentrations of chlorine. To measure a concentration outside the measurement range involved performing a dilution. The HI96771 has a unique chemistry that allows for the measurement of samples with chlorine concentrations up to 500 mg/L (ppm) without having to perform a dilution. This portable photometer features an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check™ feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration. The exclusive cuvette locking system ensures that the cuvette is inserted into the measurement cell in the same position every time to maintain a consistent path length.

Features at-a-glance

*Reagents ordered separately

cal check standards

CAL Check™ standards with certificate

The HI96771-11 CAL Check™standards are used for calibration and performance verification of photometers with the CAL Check function.

Supplied with Certificate of Analysis

  • Lot number
  • Expiration date
  • Standard value @ 25 ℃
  • Reference meter NIST traceable

Provided storage containers

  • Light tight
  • Protects from accidental breakage

Additional information

Weight 0.84 kg
Parameter

Chlorine

Specifications

SKU HI96771-web
Name Ultra High Range Chlorine Portable Photometer – HI96771
Free Chlorine Range Free Chlorine (P1): 0.00 to 5.00 mg/L (ppm)
Free Chlorine Resolution Free Chlorine (P1): 0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L
Free Chlorine Accuracy Free Chlorine (P1): ±0.03 mg/L ±3% of reading
Total Chlorine Measurement Type Ultra High Range Chlorine (P2): 1 mg/L from 0 to 200 mg/L; 10 mg/L above 200 mg/L
Total Chlorine Range Ultra High Range Chlorine (P2): 0 to 500 mg/L (ppm)
Total Chlorine Accuracy Ultra High Range Chlorine (P2): ±3 mg/L ±3% of reading
Chlorine Method adaptation of the Standard Methods for the Examination of Water and Wastewater, 4500-Cl method
Photometer/Colorimeter Light Source tungsten lamp
Photometer/Colorimeter Light Detector silicon photocell with narrow band interference filter @ 525 nm
Battery Type/Life 9V battery
Automatic Shut-Off after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder
Environment 0 to 50°C (32 to 122°F); RH max 95% non-condensing
Weight 360g (12.7oz.)
Dimensions 193 x 104 x 69 mm (7.6 x 4.1 x 2.7”)

Ordering Info

HI96771 is supplied with sample cuvettes (2) with caps, 9V battery, instrument quality certificate and instruction manual.

Manuals

Ultra High Range Chlorine Portable Photometer – HI96771