Wastewater Multiparameter (with COD) Benchtop Photometer and pH meter - HI83314

$1,694.00

In stock

HI83314 is a compact, multiparameter photometer for measuring key water and wastewater quality parameters. The meter is one of the most advanced photometers available with an innovative optical design that utilizes a reference detector and focusing lens to eliminate errors from changes in the light source and from imperfections in the glass cuvette. 10 key water and wastewater quality parameters with 20 different methods covering multiple ranges are programmed into the meter. Wastewater treatment digestion parameters include COD, Total Nitrogen and Total Phosphorous which are important for monitoring nutrient removal. The HI83314 also offers an absorbance measurement mode for performance verification and for users that would like to develop their own concentration versus absorbance curves.

To save valuable laboratory benchtop space, the HI83314 doubles as a professional pH meter with its digital pH/temperature electrode input. Now one meter can be used for both photometric and pH measurements.

Advanced optical system

Unparalleled performance from a benchtop photometer

Digital pH electrode input

Save valuable bench space with one meter that works both as a photometer and as a laboratory pH meter

Wastewater Treatment Digestion Parameters

Measures common wastewater parameters including COD, Total Nitrogen and Total Phosphorus

view compatible reagents

Description

HI83314 benchtop photometer measures 10 different key wastewater quality parameters using 20 different methods that allow for multiple ranges and variations in chemistry for specific applications. The Chemical Oxygen Demand (COD) parameter is included for industrial and municipal wastewater treatment. The Phosphorous and Nitrogen parameters included are beneficial to municipal wastewater treatment customers that need to monitor their biological and chemical nutrient removal process. This photometer features an innovative optical system that uses LED’s, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.

A digital pH electrode input is provided allowing the user to measure pH by a traditional glass electrode. The digital pH electrode has a built in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.

Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.

The HI83314 offers an absorbance measuring mode that allows for CAL Check standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the 4 wavelengths of light (420 nm, 466 nm, 525 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for education to teach the concept of absorbance by using the Beer-Lambert Law.

Features/Benefits:

Backlit 128 x 64 Pixel Graphic LCD Display

  • Backlit graphic display allows for easy viewing in low light conditions
  • The 128 x 64 Pixel LCD allows for a simplified user interface with virtual keys  and on-screen help to guide the user through use of the meter

Built-in Reaction Timer for Photometric Measurements

  • The measurement is taken after the countdown timer expires.
  • Countdown timer ensures that all readings are taken at the appropriate reaction intervals regardless of user for better consistency in measurements

Absorbance mode

  • Hanna’s exclusive CAL Check cuvettes for validation of light source and detector
  • Allows for the user to plot concentration versus absorbance for a specific wavelength for use with user supplied chemistry or for teaching principles of photometry

Units of Measure

  • Appropriate unit of measure along with chemical form is displayed along with reading

Result Conversion

  • Automatically convert readings to other chemical forms with the touch of a button

Cuvette Cover

  • Aids in preventing stray light from affecting measurements

Digital pH Electrode Input

  • Measure pH and temperature with a single probe
  • Good Laboratory Practice (GLP) to track calibration information including date, time, buffers used, offset and slope for traceability
  • pH CAL Check alerts user to potential problems during the calibration process
  • Space saving having a pH meter and photometer built into one meter

Data Logging

  • Up to 1000 photometric and pH readings can be stored by simply pressing the dedicated LOG button. Logged readings are just as easily recalled by pressing the RCL button
  • Sample ID and User ID information can be added to a logged reading using alphanumeric keypad

Connectivity

  • Logged readings can be quickly and easily transferred to a flash drive using the USB-A host port or to a computer using the micro USB-B port
  • Data is exported as a .CSV file for use with common spreadsheet programs

Battery Status Indicator

  • Indicates the amount of battery life left

Error Messages

  • Photometric error messages include no cap, high zero, and standard too low
  • pH calibration messages include clean electrode, check buffer and check probe

On-screen Features

method selection

Method Selection
Users can easily select any one of the 20 measurement methods and 4 absorbance methods via the dedicated METHOD button.

data logging

Data Logging
Up to 1000 measurement readings can be logged with user and sample ID and recalled for future use.

ph measurement mode

pH Measurement Mode
Selecting the pH measurement mode allows for the photometer to be used as a professional pH meter with many features including temperature compensated measurements, automatic two point calibration, and GLP.

Digestion Parameters Features

HI83399 cuvette adapter

Cuvette Adapter
The HI83314 is supplied with a 16 mm cuvette adapter that accepts digestion vials.

COD reagents

Digestion Vial Methods
Compatible with COD (EPA, ISO, and mercury free methods), Nitrogen and Phosphorous reagetns packaged in 16 mm digestion vial. Reagents are sold separately.

HI839800 COD reactor

COD Reactor for Digestion Vials
A COD reactor is used to heat the digestion vials. The digestion vials must be heated to a specific temperature for a period time making the HI839800 an important accessory required to have a complete wastewater treatment monitoring system.

Advanced Optical System

HI83314 is designed with an innovative optical system that incorporates a beam splitter so that light can be used for absorbance readings and for a reference detector. The reference detector monitors the intensity of light and modulates when there is drift due to power fluctuation or the heating of the optical components. Each part has an important role in providing unparalleled performance from a photometer.

photometer multiparameter optical

High Efficiency LED Light Source

An LED light source offers superior performance as compared to a tungsten lamp. LEDs have a much higher luminous efficiency, providing more light while using less power. They also produce very little heat, which could otherwise affect the optical components an electronic stability. LEDs are available in a wide array of wavelengths, whereas tungsten lamps are supposed to be white light (all wavelengths of visible light) but actually have a poor blue/violet light output.

High Quality Narrow Band Interference Filters

The narrow band interference filter not only ensure greater wavelength accuracy (± 1 nm) but are extremely efficient. The filters used allow up to 95% of the light from the LED to be transmitted as compared to other filters that are only 75% efficient. The higher efficiency allows for a brighter, stronger light source. The end result is higher measurement stability and less wavelength error.

Reference Detector for a Stable Light Source

A beam splitter is used as part of the internal reference system of the HI83314 photometer. The reference detector compensates for any drift due to power fluctuations or ambient temperature changes. Now you can rely on a stable source of light between your blank (zero) measurement and sample measurement.

Large Cuvette Size

The sample cell of the HI83314 fits a round, glass cuvette with a 25 mm path length. Along with the advanced optical components, the larger size of the cuvette greatly reduces errors in rotation from the indexing mark of the cuvettes. The relatively long path length of the sample cuvette allows the light to pass through more of the sample solution, ensuring accurate measurements even in low absorbance samples.

Focusing Lens for Greater Light Yield

Adding a focusing lens to the optical path allows for the collection of all of the light that exits the cuvette and focusing the light on the silicon photo detector. This novel approach to photometric measurements cancels the errors from imperfections and scratches present in the glass cuvette eliminating the need to index the cuvette.

Additional information

Weight 2.4 kg
Parameter

Ammonia, Chemical Oxygen Demand, Chlorine, Nitrate, Nitrite, Nitrogen, pH, Phosphorus

pH Accuracy

Photometer: ±0.1 pH pH electrode: ±0.01 pH

pH Calibration

10.01), 6.86, 7.01, 9.18, Automatic one or two point calibration with one set of standard buffers available (4.01

pH Range

Photometer: 6.5 to 8.5 pH pH electrode: -2.00 to 16.00 pH

pH Resolution

Photometer: 0.1 pH pH electrode: 0.01 pH

Specifications

SKU HI83314-01
Name Wastewater Multiparameter (with COD) Benchtop Photometer and pH meter – HI83314
pH Range Photometer: 6.5 to 8.5 pH
pH electrode: -2.00 to 16.00 pH
pH Resolution Photometer: 0.1 pH
pH electrode: 0.01 pH
pH Accuracy Photometer: ±0.1 pH
pH electrode: ±0.01 pH
pH Calibration Automatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01)
pH Temperature Compensation Automatic (-5.0 to 100.0 ℃; 23.0 to 212.0 °F); limits reduced based on the pH electrode used
pH CAL Check (electrode diagnostics) clean electrode and check buffer/check probe displayed during calibration
pH Method Photometer: phenol red
pH-mV Range ±1000 mV
pH-mV Resolution 0.1 mV
pH-mV Accuracy ±0.2 mV
Absorbance Range 0.000 to 4.000 Abs
Absorbance Resolution 0.001 Abs
Absorbance Accuracy ±0.003Abs @ 1.000 Abs
Ammonia Range Low Range: 0.00 to 3.00 mg/L
Low Range (16 mm vial): 0.00 to 3.00 mg/L
Medium Range: 0.00 to 10.00 mg/L
High Range: 0.0 to 100.0 mg/L
High Range (16 mm vial): 0.0 to 100.0 mg/L
(all as NH3-N)
Ammonia Resolution 0.01 mg/L; 0.1 mg/L
Ammonia Accuracy Low Range: ±0.04 mg/L ±4% of reading
Low Range (16 mm vial): ±0.10 mg/L or ±5% of reading
Medium Range: ±0.05 mg/L ±5% of reading
High range: ±0.5 mg/L ±5% of reading
High range (16 mm vial): ±1 mg/L or ±5% of reading
Ammonia Method Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method
Chemical Oxygen Demand Range Low Range: 0 to 150 mg/L
Medium Range: 0 to 1500 mg/L
High Range:0 to 15000 mg/L
Chemical Oxygen Demand Resolution Low and Medium Range: 1 mg/L
High Range: 10 mg/L
Chemical Oxygen Demand Accuracy Low Range: ±5 mg/L ±5% of reading
Medium Range: ±15 mg/L ±4% of reading
High Range: ±150 mg/L ±3% of reading
Chemical Oxygen Demand Method Adaptation of the USEPA 410.4
ISO dichromate methods
Mercury-free dichromate green method (LR & MR); dichromate method (HR)
Total Chlorine Range 0.00 to 5.00 mg/L (as Cl2)
Total Chlorine Resolution 0.01 mg/L
Total Chlorine Accuracy ±0.03 mg/L ±3% of reading
Chlorine Method Adaptation of the EPA 330.5 DPD method
Nitrate Range 16 mm vial: 0.0 to 30.0 mg/L (as NO3– N)
Nitrate Resolution 0.1 mg/L
Nitrate Accuracy 16 mm vial: ±1.0 mg/L or ±3% of reading
Nitrate Method Adaptation of the cadmium reduction method
Nitrite Range Freshwater
Low Range: 0 to 600 µg/L (as NO2-N)
High Range: 0 to 150 mg/L (as NO2)
Nitrite Resolution Freshwater: 1 µg/L; 1 mg/L
Nitrite Accuracy Freshwater
Low Range: ±20 µg/L ±4% of reading
High Range: ±4 mg/L ±4% of reading
Nitrite Method Adaptation of the EPA Diazotization method 354.1
Nitrogen, Total Range Low Range: 0.0 to 25.0 mg/L (as NO3-N)
High Range: 0 to 150 mg/L (as N)
Nitrogen, Total Resolution 0.1 mg/L; 1 mg/L
Nitrogen, Total Accuracy Low Range: ±1.0 mg/L or ±5% of reading
High Range: ±3 mg/L or ±4% of reading
Nitrogen, Total Method Chromotropic acid method
Phosphorous, Acid Hydrolyzable Range 0.00 to 1.60 mg/L (as P)
Phosphorous, Acid Hydrolyzable Resolution 0.01 mg/L
Phosphorous, Acid Hydrolyzable Accuracy ±0.05 mg/L or ±5% of reading
Phosphorous, Acid Hydrolyzable Method Adaptation of the EPA method 365.2 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 4500-P E, ascorbic acid method
Phosphorous, Reactive Range Low Range: 0.00 to 1.60 mg/L (as P)
High Range: 0.0 to 32.6 mg/L (as P)
Phosphorous, Reactive Resolution 0.01 mg/L; 0.1 mg/L
Phosphorous, Reactive Accuracy Low Range: ±0.05 mg/L or ±4% of reading
High Range: ±0.5 mg/L or ±4% of reading
Phosphorous, Reactive Method Adaptation of Standard Methods for the Examination of Water and Wastewater, 20th edition and EPA method 365.2 (low range)
Low Range: ascorbic acid method
High Range: vanadomolybdophosphoric acid method
Phosphorous, Total Range Low Range: 0.00 to 1.15 mg/L (as P)
High Range: 0.0 to 32.6 mg/L (as P)
Phosphorous, Total Accuracy Low Range: ±0.05 mg/L or ±6% of reading
High Range: ±0.5 mg/L or ±5% of reading
Phosphorous, Total Resolution Low Range: 0.01 mg/L; 0.1 mg/L
Phosphorous, Total Method Adaptation of Standard Methods for the Examination of Water and Wastewater, 20th edition and EPA method 365.2 (low range)
Low Range: ascorbic acid method
High Range: vanadomolybdophosphoric acid method
Input Channels 1 pH electrode input and 5 photometer wavelengths
pH Electrode digital pH electrode (not included)
Photometer/Colorimeter Light Source 5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters
Photometer/Colorimeter Light Detector silicon photodetector
Bandpass Filter Bandwidth 8 nm
Bandpass Filter Wavelength Accuracy ±1 nm
Cuvette Type round, 24.6 mm
Number of Methods 128 max.
GLP calibration data for connected pH electrode
Display 128 x 64 pixel LCD with backlight
Logging Type log on demand with user name and sample ID optional input
Logging Memory 1000 readings
Connectivity USB-A host for flash drive; micro-USB-B for power and computer connectivity
Power Supply 5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included)
Battery Type/Life 3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement
Environment 0 to 50.0 ℃ (32 to 122.0 °F); 0 to 95% RH, non-condensing
Weight 1.0 kg (2.2 lbs.)
Dimensions 206 x 177 x 97 mm (8.1 x 7.0 x 3.8 in.)